View current Department of Defense opportunities →Inductively Coupled Plasma - Optical Emission Spectrometry
AI Overview
The Army is seeking a new benchtop ICP-OES instrument for simultaneous analysis of aqueous and organic samples, with a 180-position autosampler, Windows 11 software, and UHP argon as the only gas input. This is a standard commercial equipment purchase for laboratory elemental analysis, with a brand-name-or-equal solicitation open to small businesses.
This summary is AI-generated from the official solicitation.
Key Details
Official Description
--- Attachment text --- [W911S227U0002_S2P2_Brand_Name_or_Equal_Pirolli__Alec_L_CIV_USARMY_RIA_JMTC__USA__2026_11_09_15_35_02__EST_.pdf] Combined Synopsis/Solicitation W911S227U0002 2026-11-09 15:34:59 (EST) This is a combined synopsis/solicitation for commercial products prepared in accordance with the format in subpart 12.6, as supplemented with additional information included in this notice. This announcement constitutes the only solicitation; proposals are being requested and a written solic...
Change History
Critical change · Requirements + Scope + Eligibility + Description
Inductively Coupled Plasma - Optical Emission Spectrometry
- Dual-view requirement removed: The mandatory dual-view configuration is gone, so radial-only ICP-OES systems can now compete alongside dual-view systems.
- Brand-name references removed: The Agilent 5800 reference and brand comparisons were deleted; the item is still posted as "brand name or equal," but the Agilent listing remains only as a header.
- 64-bit software constraint dropped: Windows 11 compatibility and automated interference correction are kept, but the 64-bit architecture requirement is removed.
- Hydrofluoric acid capability made optional: HF-resistant sample introduction hardware is now a separate line item or manufacturer kit option rather than base hardware.
- Gas requirement added: Ultra-High Purity Argon is the only required gas; systems needing a secondary purge gas such as nitrogen are not acceptable.
- Cooling flexibility: A separate water chiller is no longer mandatory; either integrated air cooling or a vendor-supplied external recirculating chiller is acceptable.
- Warranty, installation, and training formalized: A one-year warranty covering parts, labor, and travel, plus certified on-site installation and operator training, are now explicit requirements.
- Performance criteria deleted: The EPA detection limit and ±10% certified metal standard accuracy requirements were removed.
Why it matters: Applicants with radial-only or lower-cost configurations can now bid, and HF capability and chiller type are no longer disqualifiers. Quotes must still include UHP Argon-only operation and the warranty, installation, and training terms.
Informational change · Status
Inductively Coupled Plasma - Optical Emission Spectrometry
The solicitation's status changed from open to closed. Applicants can no longer submit proposals through this opportunity, so any work on an application for it should stop. If you were preparing a submission, check whether the agency has posted a successor announcement or an amendment reopening intake, and watch the posting for any note on late submissions or a reopening date before deciding whether to pursue other funding routes.
Inductively Coupled Plasma - Optical Emission Spectrometry
New opportunity: Inductively Coupled Plasma - Optical Emission Spectrometry
Get Alerts for Opportunities Like This